One-Port VNA Calibrations

How do I perform VNA calibrations with my Pyramid ProbeTM cards?

For one-port VNA calibrations, use short-open-load calibration coefficients for an equivalent Air Coplanar® (ACP) probe pitch and configuration (ground-signal-ground or ground-signal). Use the impedance standard substrate indicated in the following table: 

Air Coplanar Probe (ACP) Calibration Coefficients
C-Open (fF) GSG
L-short (pH)
 
L-Term (pH) GSG ISS P/N Probe Pitch (um) GS/SG ISS P/N C-Open (fF) GS/SG
L-short (pH)
 
L-Term (pH) 
-9.32.4-3.5101-190100103-726-11.033.536.5
-9.53.6-2.6101-190125103-726-11.041.747.2
-9.74.8-1.7101-190150103-726-11.049.857.8
-10.17.20.2101-190200103-726-11.066.279.2
-10.59.62.1101-190250103-726-11.082.5100.5
-15.711.0-25.0106-682250106-683-7.027.00.0
-13.615.8-21.0106-682350106-683-7.028.20.0
-12.618.2-19.0106-682400106-683-7.028.80.0
-10.523.0-15.0106-682500106-683-7.030.00.0
-9.628.1-3.3106-682650106-683-6.442.914.1
-931.64.4106-682750106-683-6.051.623.4
-7.540.423.6106-6821000106-683-5.073.446.6
-649.142.9106-6821250106-683-4.095.169.9

Coefficients for VNA calibration depend on the style and pitch of the probe, as well as the ISS used. Calibration coefficients are also suitable for corresponding Pyramid Probe contact configurations.